Volumes  Volume 3  Contents of Current Session  
Adv. Radio Sci., 3, 51-58, 2005
www.adv-radio-sci.net/3/51/2005/
doi:10.5194/ars-3-51-2005
© Author(s) 2005. This work is licensed
under a Creative Commons License.


Influence of nonideal LRL or TRL calibration elements on VNA S-parameter measurements

U. Stumper
Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany

Abstract. For the 7-term LRL and TRL calibration of a four-sampler vector network analyser (VNA), expressions for the deviations of the measured S-parameters of two-port test objects from their actual values are presented as functions of the deviations of the S-parameters of the LRL/TRL calibration elements from their ideal values. The obtained sensitivity coefficients are suitable for establishing the Type-B uncertainty budget for S-parameter measurements. They show how the measurements are affected by imperfect calibration elements and nonideal connections.

Full Article in PDF (PDF, 122 KB)   

Citation: Stumper, U.: Influence of nonideal LRL or TRL calibration elements on VNA S-parameter measurements, Adv. Radio Sci., 3, 51-58, doi:10.5194/ars-3-51-2005, 2005.   Bibtex   EndNote   Reference Manager    XML
 

Recent Papers